应用领域:环境/能源,环境水/废水/饮用水
检测样品:溶剂残留
检测项目:分离与检测
参考标准:无
Column: NanoChrom BP-624
Dimension: 30m × 0.32mm × 1.80um
Carrier: Hydrogen, flow 2.3 mL/min
Inlet: Split, 260 °C, split column flow: 60 mL/min
Oven: 40 °C (hold 20min) to 240 °C (hold 20min) at 10 °C /min
Sample: USP class 2 residual solvent mixture A
Detector: FID 280 °C