福禄克 5960A Triple Point of Argon System
价格:面议

福禄克 5960A Triple Point of Argon System

产品属性

  • 品牌福禄克
  • 产地北京
  • 型号 5960A Triple Point of Argon System
  • 关注度281
  • 信息完整度
  • 供应商性质生产商
  • 产地类别国产
关闭
产品描述

Breakthrough primary temperature standards capability

  • Low uncertainty of 0.25 mK

  • Temperature plateau duration longer than 30 hours

  • Four entrant wells with radial uniformity less than 0.05 mK

  • 480 mm total immersion depth

  • Argon purity of 99.9999 %

  • Multi-language display support

The Fluke Calibration 5960A Triple Point of Argon System enables primary standards laboratories to realize the argon triple point using a consistent, reproducible process with the lowest uncertainty of any commercially available system. The Fluke Calibration system produces a 30-hour temperature plateau and provides multiple entrant wells to increase SPRT calibration efficiency and throughput.

Low calibration uncertainty

A number of technology advancements are designed into the Fluke Calibration 5960A system to achieve a low uncertainty of 0.25 mK. For example, the argon cell is carefully built with an ultra-clean process using 99.9999 % (six nines) purity argon resulting in an uncertainty factor from impurities of 0.015 mK. The deep, 480 mm immersion depth of the entrant well minimizes the effects of stem conduction to less than 0.01 mK and radial uniformity to less than 0.05 mK.

High lab throughput

The 5960A system utilizes liquid nitrogen (LN2) in a Dewar vessel surrounding an argon cell to achieve –189.3442 °C. A digital controller combined with a heater maintains a temperature plateau lasting up to 30 hours with 0.1 mK variation without using complicated pressure controls. By regular renewal of liquid nitrogen in the system, the plateau length can even be extended to four days. The system’s long, stable temperature plateau lets you carry out your SPRT calibrations with better quality and more efficiency. Four entrant wells increase lab productivity by allowing you to calibrate four SPRTs simultaneously.

Ease of operation and maintenance

The display panel of the 5690A provides a simple menu structure with soft function keys for setup and programming the argon system. The readout displays simultaneously the set-point temperature, actual control sensor temperature and percent heater power. The display is selectable in eight languages—English, French, Spanish, Italian, German, Russian, Chinese, and Japanese. Applying positive pressure with helium or dry nitrogen to an entrant well port prevents your SPRTs from being bound in the entrant well by freezing moisture. You can insert and remove SPRTs as desired from the system over the duration of a single plateau. Other systems require that you first melt the cell in order to add or remove an SPRT.

Specifications
Assigned value-189.3442 °C
Argon gas purity99.9999 % (6N purity)
Uncertainty (k=2)0.25 mK
Length of plateau (0.1 mK)>30 hours
Argon cell depth*160 mm
Number of entrant wells4
Radial uniformity0.05 mK
Total immersion depth480 mm
Entrant well inner diameter (ID)8.0 mm
Argon gas volume13.4 moles (535 grams)
Liquid nitrogen Dewar volume44.2 liters
Display resolution0.001 °C
Set-point accuracy0.1 °C (adjustable)
Power requirements100 V to 115 V (± 10 %) 50/60 Hz, 230 W 230 V (± 10 %) 50/60 Hz, 230 W
System fuse ratings 115 V: 2 A T 250 V
230 V: 1 A T 250 V 
Size (HxWxD) 952 mm x 673 mm x 483 mm (37.5 in x 26.5 in x 19 in) 
Weight 94 kg (207 lbs) 
*Measured from the bottom of the probe well to the surface of argon sample
配件描述
5960ACC-PUMP/DEWAR

Pump and Dewar Accessory Kit 
Includes:

  • Rotary Vane Vacuum Pump

  • Self-pressurized Liquid Nitrogen Dewar (50 Liters)

  • Hoses

  • Fittings

  • Associated Hardware

5960ACC-PUMP

Pump Accessory Kit 
Includes:

  • Rotary Vane Vacuum Pump

  • Hose

  • Associated Hardware

5960ACC-DEWAR

Dewar Kit
Includes:

  • Self-Pressurized Liquid Nitrogen Dewar (50 Liters)

  • Hoses

  • Fittings

  • Associated Hardware