Calcium Flux Measurement Using Indo-1

2019.4.23
头像

184****5725

致力于为分析测试行业奉献终身

实验概要

Measure the calcium in cytosol with indo-1.

主要试剂

DMSO, Imonomycin, RPMI1640, PBS

实验步骤

1. Dissolve the contents of 1 vial (50 μg indo-1) in 50 μl dry DMSO. This makes a 1mM solution. T and B cells can be loaded with 1 μM indo-1; granulocytes or monocytes with 1-3 μM and dendritic cells with 3-6 μM Indo-1. Cell concentration on loading should be 1-10 x 106/ml.
2. Incubate cells at 37°Cfor 45-60 min in a waterbath or incubator. Wash twice in serum free medium or PBS. Resuspend at 106/ml in PBS or serum free medium of choice. Either keep the cells at 37°Cif using immediately or keep on ice until ready for use; prior warming back to RT or 37°Cshould be done as Indo-1 fluorescence is temperature dependent.
3. Immunophenotyping if required should then be carried out on ice.
4. Cells can then be analysed on BD LSR which allows up to six-colour analysis. Thus FITC, PE, TC and APC can be used for immunophenotyping. Indo-1 is only loaded into viable cells thus there is no requirement to exclude dead cells by PI.
5. The filter set-up on the BD LSR for Indo-1 (UV excitation only) is for calcium bound Indo-1 violet FL-5 424/44 nm BF filter and unbound Indo-1 green FL-4 530/30nm BF filter. Calcium flux is measured as a Ratio between calcium bound Indo-1 and unbound or FL-5/FL-4 versus time.
6. Full scale deflection of the calcium flux is measured by the addition of ionomycin 1-10 μg/ml. Contribution of internal stores of calcium can be measured by resuspending cells in calcium-free medium and then adding ionomycin.