技术规格
Specifications | Dansensor MultiCheck 2 | |
Test types | Headspace gas analysis; leak detection (burst and leak options); optional valve leak and valve release tests | |
Typical test time | 60-90 seconds (combined headspace and leak test on 500 gram bag); dependent on test settings, package size, etc. | |
Heating time (max) | 10 min | |
Applications | Coffee (ground and whole bean) bags and packaged pillow packs | |
Package size (max) | 5 liters | |
Headspace analysis | Leak detection | |
Sample volume (typical) | 25.7 ml | n/a |
Measuring range | 0-100% O2 | 10-1000 mbar (0.15-14.5 psi) |
Accuracy of measurement | ±0.25% O2+ 2% of reading | Low range: ±0.5 mbar (0.007 psi) or ± 0.5% of reading |
Number of test definitions | 999 | |
Data collection capacity | Test data from more than 100,000 tests | |
Data export | Stored data to CSV file (USB) or real time data transfer CSV-string per test (RJ-45) | |
Dimensions (WxDxH) and weight | Instrument: 304 x 236.5 x 185 mm (12” x 9.3” x 7.3”) 4.4 kg (9.7 lbs) | |
Calibration and service | 12 months | |
Ambient temperature | Operational: +2°C to +35°C Storage: -20°C to +60°C | |
Relative humidity | Operational: 10 to 90 %RH (non-condensing) | |
Ambient pressure | Operational: 750-1100 mbar | |
Power supply and consumption | External PSU, 100-240 VAC, 47-63 Hz | |
Air supply pressure and connection | 2-5 bar Ø6/4mm tube DIN ISO 8573-1:2010 [4:4:3] | |
Air consumption | Max 5 L/min (depending on test settings) | |
Connections | LAN: RJ-45 Ethernet 10/100 Mbit/s, DHCP client or fixed IP USB: 1 x Host, USB 2.0 Type A; 1 x Device, USB 2.0 Type B (max current 500 mA) | |
Consumables and accessories | Septa, VTU (valve test unit) | |
Warranty | 12 months | |
Compliance | CE, China RoHS II | |
Standards | ASTM F2095, ISO 11607 |
可集成阀门测试单元 (VTU)
带 USB 连接器的条码扫描器
带 USB 连接器的键盘
隔垫
PTPS AMECare 计划