• compact turn-key tunable Ti:Sapphire femtosecond laser
laser pulse width: 200 fs
repetition frequency: 80 MHz
in situ laser power: 2-50 mW
wavelength range. 710-920 nm
• full-frame scanning, region-of-interest (ROI) scanning,
line scanning, single-point illumination (spot scan)
• typical scan range: 350 μm x 350 μm (horizontal); 200 μm (vertical)
• spatial resolution: < 0.5 μm (horizontal); < 2 μm (vertical)
• focusing optics: magnification 40x NA 1.3
• control and image processing software (JenLab Scan, JenLab Image)
• operating temperature 15-35°C
• relative humidity: 5-65 %
• power requirements: 230 VAC (50 Hz) or 115 VAC (60 Hz)
• CE certified class 1M /IIa medical product
• 710 mm x 960 mm x 1400 mm, 250 kg
The system requires an air-conditioned room with reduced ambient light.
Notes: These specifications are subject to change without notice.