规 格:
载荷容量:20KN、50KN、100KN、250KN、300KN
载荷精度:显示值的±0.5%
(保证范围:载荷传感器容量的1/1~1/1000。100KN以下机型)
载荷量程:一个量程不需要转换
试验速度:0.0005~1000mm/min
采样间隔:0.2msec
十字头速度精度:0.1%
Floor Test Frame立式 | 立式 | 立式 | ||
---|---|---|---|---|
Model Type and Name | AG-X Plus 50 kN | AG-X Plus 100 kN | AG-X Plus 300 kN | |
Test Frame Load Capacity | Max. 20/50 kN (4,400/11,200 lbs) | Max. 100 kN (22,400 lbs) | Max. 300 kN (67,000 lbs) | |
Crosshead Speed Range | 0.0005 to 1000 mm/min (0.00002 to 39 in/min) | 0.0005 to 500 mm/min (0.00002 to 19 in/min) | ||
Maximum Return Speed | 1200 mm/min (47 in/min) | 600 mm/min (23 in/min) | ||
Crosshead Speed Accuracy | Within ±0.1% of test speed | |||
Crosshead Positional Accuracy | ±0.1% of indicated value or ±0.01mm, whichever is larger | |||
Test Force Measurement Accuracy | High-Accuracy Load Cells | Within ± 0.5% of indicated test force (for 1/1 to 1/250 of load cell rated capacity) Conforms to ISO 7500-1 Class 0.5, ASTM E4, JIS B7721 Class 0.5, EN 10002-2 Grade 0.5, BS1610 Class 0.5, and DIN51221 Class 1 | ||
Standard-Accuracy Load Cells | Within ± 1% of indicated test force (for 1/1 to 1/1000 of load cell rated capacity) Conforms to ISO 7500-1 Class 1, ASTM E4, JIS B7721 Class 1, EN 10002-2 Grade 1, BS1610 Class 1, and DIN51221 Class 1 | |||
Within ± 1% of indicated test force (for 1/1 to 1/500 of load cell rated capacity) Conforms to ISO 7500-1 Class 1, ASTM E4, JIS B7721 Class 1, EN 10002-2 Grade 1, BS1610 Class 1, and DIN51221 Class 1 | ||||
Effective Width | 600 mm (23.5 in) | |||
Crosshead-Travel for Standard Height Models | Max. 1265 mm (50 in) | Max. 1250 mm (49 in) | Max. 1440 mm (57 in) | |
Crosshead-Travel for +250mm Extended Models | Max. 1515 mm (60 in) | Max. 1500 mm (59 in) | Max. 1690 mm (67 in) | |
Crosshead-Travel for +500mm Extended Models | Max. 1765 mm (69 in) | Max. 1750 mm (69 in) | Max. 1940 mm (76 in) | |
Crosshead-Travel for +750mm Extended Models | Max. 2015 mm (79 in) | Max. 2000 mm (79 in) | Max. 2190 mm (86 in) | |
Data Capture Rate | 5000 Hz | |||
Frame Rigidity | Min. 175 kN/mm (1548 lbs/in) | Min. 300 kN/mm (2655 lbs/in) | Min. 400 kN/mm (3540 lbs/in) | |
Standard Input/Output Channels | –1 dedicated load channel –2 analog output channels –2 analog input channels –2 digital input channels –3 internal amplifier card slots (one is used for analog input channels*4) | |||
Dimensions for Standard Height Models | W1186 x D752 x H2173 mm (W47 x D30 x H86 in) | W1186 x D752 x H2423 mm (W47 x D30 x H95 in) | ||
Weight for Standard Height Models | Approx. 654 kg (1,441 lbs) | Approx. 834 kg (1,838 lbs) | Approx. 960 kg (2,116 lbs) | |
Operating Environment | Temperature: 5° C to 50° C (41° F to 122° F); Humidity: 20% to 80% (no condensation) Floor Vibrations: frequency 10 Hz max., amplitude 5 µm (0.0002 in) max. | |||
Power Requirements | Three Phase 200-230 V 50/60 Hz |
Note 1: When the load cell capacity is smaller than the tester load capacity, the former is the maximum test force.
Note 2: Shimadzu recommends validation at an installation site that meets the requirements specified in these standards.
Note 3: Analog input card can be swapped for another amplifier card.
Note 4: Wide frame models (975 mm, 1100mm, and 1375mm), high capacity models (500kN, 600kN, 1000kN, and 2000kN), and ultralow/high-speed models are available upon request .
Operating System | Windows 7 | |
Type | 32 or 64 bit | |
CPU | Intel Pentium 4, 2GHz or higher | |
Memory | 1 GB min. | |
HDD | 80 GB or greater | |
CD-ROM Drive | Required for software installation | |
USB Port | 1 port for main frame, other accessories may require additonal ports | |
Display | Size | 19 inch minimum |
Resolution | 1024 x 768 pixels minimum |