Fluorolog-QM系列超灵...
参数指标我要纠错



PTI QuantaMaster System Specifications


PTI QuantaMaster 8000
Series
PTI QuantaMaster 300 SeriesPTI QuantaMaster 800
Signal to Noise Ratio30,000:1 or better300: 1500:1 or better
300 Plus: 3000:1 or better
6000:1 or better
Data Acquisition Rate1,000,000 points/sec. to
1 point/1000 sec
300/300 Plus: 1 point/sec to
300 points/sec
310: Up to 20 points/sec
1,000,000 points/sec.
to 1 point/1000 sec
Inputs4 photon counting (TTL); 4 analog (+/- 10 volts) ; 1 analog reference channel (+/- 10 volts); 2 TTL
Outputs2 analog (+/- 10 volts) 2 TTL
Emission Range185 nm to 900 nm
(optional to 5,500 nm)
185 to 680 nm
(optional to 5,500 nm)
185 to 680 nm
(optional to 1,010 nm)
Light SourceHigh efficiency “ECO” friendly
continuous 75 W Xenon arc
lamp (Optional 450 W xenon)
300/300 plus: High power
Xenon flash lamp
310: Nitrogen laser pumped
dye laser
High efficiency “ECO” friendly
continuous 75 W Xenon arc
lamp
Monochromator300 mm, coma-aberration corrected, asymmetrical, excitation or emission optimized, Czerny-Turner design
Focal Length300 mm300 mm150 mm excitation
300 mm emission
SlitsComputer controlled,
continuously adjustable
Manual or computer
controlled, continuously
adjustable
Manual continues Excitation,
Manual or computer controlled
continuous Emission
Emission Grating1200 line/mm 400 nm blaze
(Plus two optional gratings)
1200 line/mm 400 nm blaze
(Plus two optional gratings)
1200 line/mm 400 nm blaze
(Plus two optional gratings)
Optional Grating
75 to 2400 line/mm and holographic models available
Wavelength
Accuracy
+/- 0.3 nm+/- 0.3 nm+/- 0.05 nm excitation
+/- 0.3 nm emission
Minimum Step Size0.022 nm0.022 nm0.05 nm excitation
0.022 nm emission
Lifetime RangeN/A1 μs to secondsN/A
DetectionMultimode: Photon Counting, 3 analog (fast, medium, slow response), direct and Single-Shot
Transient Digitizer (SSTD) mode
System ControlComputer interface with spectroscopy software




相关分子荧光光谱/荧光分光光度计