Contents | EMIA-Expert | |
---|---|---|
Required sample amount | 1g ± 0.1g | |
Typical Meas. time | 70 seconds/cycle | |
Carbon | Meas. Range (m/m) | 0.6 ppm - 10.0% |
Blank accuracy | σn-1≦ 0.3 ppm | |
Sulfur | Meas. Range (m/m) | 0.6 ppm - 1.0% |
Blank accuracy | σn-1 ≦ 0.3 ppm | |
Sample meas. Accuracy | Carbon | σn-1 ≦ 0.3 ppm(less than 20 ppm) σn-1 ≦ 1.0 ppm or RSD≦0.5%(more than 20 ppm) |
Sulfur | σn-1 ≦ 0.3 ppm(less than 20 ppm) σn-1 ≦ 1.0 ppm or RSD≦0.75%(more than 20 ppm) | |
Utility | Carrier gas | Oxygen (Purity; 99.5%, Pressure; 0.3-0.33 MPa) |
Operation gas | Nitrogen (Purity; 99.5%, Pressure; 0.35-0.38 MPa) | |
Power | 200/220/240V, 50/60Hz, 5kVA | |
Dimensions | Main unit* | 500(W) x 725(D) x 710(H) mm |
Purifier unit | 160(W) x 560(D) x 512(H) mm | |
Mass | Main unit | approximately 134 kg |
Purifier unit | approximately 21 kg | |
Data processing and operation | USB data communication with PC Windows 8.1 Touch panel or Key board and mouse operation | |
Model name | EMIA-20E: Carbon/Sulfur EMIA-21E: Carbon EMIA-22E: Sulfur |
Note) “ppm“ is equal to mg/kg.
*Except piping parts, projection parts (D:79mm) and flow meter on the back side.