EMGA-920 氧/氮分析仪

EMGA-920 氧/氮分析仪参数指标

参数指标我要纠错

  EMGA-920:氧/氮分析仪

  原理

  氧气:非分散红外探测器

  氮:热导检测器(TCD)

  测量范围*:氧气:0-5%,氮气:0-3%

  *通过降低样品重量可以达到100%的重量。

  样品重量:1.0±0.1g

  灵敏度(zei小读数):氧/氮:0.0000001%(m / m)

  精度**(重复):

  氧气/氮气

  σn-1≤0.000002%(m / m)或CV≤0.5%(以较大者为准)(参考气体)

  σn-1≤0.00003%(m / m)或CV≤1.0%(以较大者为准)(标准样品)

  **根据样品重量为1.0±0.1g

規格

EMGA-920:Oxygen/Nitrogen analyzer

Principle
Oxygen:Non Dispersive Infrared detector
Nitrogen:Thermal Conductivity detector(TCD)

Measurement range*:
Oxygen:0-5%, Nitrogen:0-3%
*Up to 100% wt is possible by decreasing the sample weight.

Sample weight: 1,0 ± 0,1g

Sensitivity(Minimum reading): Oxygen/Nitrogen: 0.0000001%(m/m)

Accuracy**(Repeatability)
Oxygen/Nitrogen
σn-1≦0.000002%(m/m) or CV≦0.5% whichever is larger (Reference gas)
σn-1≦0.00003%(m/m) or CV≦1.0%  whichever is larger (Standard sample)
** Based on the sample weight 1,0 ± 0,1g and with automation functions described below

Display
1) Measurement result: LCD or printout
2) Alarm message: LCD or printout
3) Flow sheet: LCD

Power of furnace
Iner gas fusion in impulse furnace
Enable to set the generating power 0 - 8,0kw

Sample introduction: Dual sample/flux introduction mechanism

Automation functions: Auto cleaner, Crucible loader

Integration conditions:
Preset time integration
Integration up to reach comparator level
Preset time integration or integration up to reach comparator level, whichever is shorter

Sample ID: Enter up to 20 one-byte characters

Calibration
1) One point or multi point calibration (Reference gas or standard samples)
2) Calibration using previous analysis data
3) Calibration curve correction function

Functions
1) Display of realtime extraction curve
2) Analysis interruption
3) Self diagnosis and alarm display
4) Analysis of extraction curve (Superposition, difference etc analysis)
5) Out put (RC-232C or TCP/IP)

Dimensions: 750mm(H)×785mm(D)×653mm(W)
Sample introduction is positioned at 650mm from table.

Weight: 230kg
On carrying, it is devided in two units (Under 140kg/unit)

Computer: PC model operating OS Windows® XP(SP2) and beyond

Power
Main unit: AC200/220/230/240V ± 10%
Auto cleaner: AC100V (Step-down transformer)
Frequency: 50/60 ± 1Hz

Electric power consumption
Main unit:12kVA (MAX)
Auto cleaner: 1,5kVA (MAX)

Ground resistance: Less than 10Ω

Installation condition
Operation temperature: 5-40°C
Performance temperature: 5-35°C
Humidity: Maximal relative humidity 80%RH between 5-31℃, Linearly decrease down to 50%RH between 31-40℃
Vibration: Duplex amplitude 20micron and less than 0.098m/S2 accelerations at frequency band

Required gas
He carrier gas: Purity greater than 99.995%, Pressure 0.35MPa
Stainless steel pipe(O.D.3mm) and suitable connecter fitting within 3m of unit
Dry air or N2 as operating gas: Pressure 0.45MPa
Nylon pipe(O.D.6mm) and suitable connecter fitting within 5m of unit

Cooling mechanism: Radiator unit (Separation type)

Electronic balance (Option)
Enable connection with electronic balance with 1-0,01mg minimum display

Optional automatic voltage regulator (AVR)
Capacity:15kVA

 

 