两级热调制器
使用一个连续的冷喷嘴和一个脉冲化的热喷嘴,将捕集到的样品重新传送
采用液氮作为冷却剂,zei低温度可以到-189度,可对非常宽范围的样品(C4~C55)进行调制。
可安装在Agilent 7890GC, 6890GC, 5973QMS, Shimadzu GC-2010 & GCMS-QP2010Plus, Thermo Trace GC上
可以与气质联用,组成GCXGC-MS,得到更多定性信息。
可以与飞行时间质谱联用,组成GCXGC-TOFMS。
Features:
2 stage loop modulator
Liquid N2 cooling bath (optional 5, 10 or 20 liters)
Modulation rated for C3+
Installs on Agilent 7890GC, 6890GC, 5973QMS, Shimadzu GC-2010 & GCMS-QP2010Plus, Thermo Trace GC
Secondary oven installed as an option
Secondary oven can lead or lag the main oven up to 40°C
No moving parts
Modulation cycle range from 1000ms – 32000ms
Benefits:
Easy to install; most users can manage own installation within 2 hours
Can be uninstalled and allow GC to be used for other methods. Not necessary to have dedicated GC
More robust than the quad jet
Sharpest known modulation gives sharpest secondary peaks, hence highest possible secondary peak capacity.
Higher sensitivity than non-thermal modulators due to peak focusing
Compatible with MS at ordinary flow rates, unlike non-thermal modulators