比较LC-MS/MS一维色谱、二维色谱和自动在线样品制备定量分析25-OH维生素​​D3/D2

应用领域:

检测样品:

检测项目:

方案摘要

前言:

  High performance liquid chromatography – tandem mass spectrometry (HPLC-MS/MS) is now widely accepted for measurement of vitamin D metabolites. Many clinical research laboratories use 1-dimensional (1D) chromatography (for example, a single HPLC pump and chromatography column) with a triple stage quadrupole mass spectrometer. Various sample cleanup protocols, such as solid phase extraction (SPE), liquid-liquid extraction (LLE), and protein precipitation (PPT), have been applied in these analyses. Frequently, interfering peaks are seen in 25-OH vitamin D3 chromatograms, adversely affecting peak integration and leading to poor accuracy and reproducibility. Here we investigate the use of 2-dimensional chromatography using TurboFlow technology to remove all interfering peaks and significantly improve data quality.

仪器:

Thermo Scientific TSQ Quantum mass spectrometer

结论:

  The TurboFlow method described here has been developed and validated to industry recommended guidelines for clinical laboratories.

  Isobaric interferences observed with a 1D LC-MS/MS method at low 25-OH D3 metabolite concentrations were much reduced by using a 2D-LC-MS/MS approach, and even further improved by using TurboFlow technology. The Transcend™ TLX-1 LC-MS/MS with TurboFlow technology improved the sensitivity and the signal-to-noise ratio.

店铺 下载
咨询留言 一键拨号