默克密理博NovaSeptumzei高无菌等级的无菌取样装置参数指标

参数指标我要纠错

规格

NovaSeptum®Sterile Sampling Units

High Purity Flat BagAutoclavable
Flat Bag
Bottle
Sampling Unit Volume50 mL to 1000 mL50 mL to 1000 mL60, 125, 250, and 500 mL
Maximum Pressure Conditions (at 25°C )0.50 bar (7.25 psi) up to 250 mL0.30 bar (4.35 psi)0.50 bar (7.25 psi)
0.30 bar for the multi-sampling unit
0.30 bar (4.35 psi) for 1000 mL and multi sampling
Temperature Range50 mL to 250 mL: -80° (when used with a NovaSeptum Case) to 50 °C (-112 to 122 °F)
1000 mL: -20° to 50 °C (-4 to 122 °F)
-20 to 125 °C (-4 to 257 °F)-80 to 50 °C (-112 to 122 °F)
Materials of Construction
Trigger  
Septum:Medical Grade platinum-cured silicone
Body:Polyester
Cannula:ASTM 316L Stainless steel
Sampling Bag:Polyethylene film (PureFlex film)Polypropylene filmPolyethylene Terephtalate Glycol
Fluid Contact Layer:Polyethylene film (PureFlex film)Polypropylene filmPolyethylene Terephtalate Glycol
Tubing:PolyethyleneSiliconePolyethylene
Drainage Bag (valid for Multi-Sampling Application):Polyethylene film (PureFlex film)Not ApplicablePolyethylene film (PureFlex film)
Fittings
Inlet Tubing:Septum with a 1 mm or a 2 mm needle 
Outlet Tubing:Tubing: 3-piece Luer, containing a male, female and an injection site Tubing: Male Luer with Female capTubing: Male Luer Fitting and Female to Female Luer
Autoclaving*Not autoclavableAutoclavableNot autoclavable
USP <88> Class VIAll components in contact with sampling liquid conform with USP <88> Class VI.
Integrity TestingUnits are integrity tested at regular intervals during manufacturing.
AssemblyAssembled under ISO Clean Room Class 8 conditions in a facility certified to ISO 14644-1
SterilizationThe sampling units are sterilized by beta irradiation at ≥ 25 kGy according to ISO 11137The sampling units are sterilized by gamma irradiation at ≥ 25 kGy according to ISO 11137
Bacterial EndotoxinAqueous extraction contains
* Autoclaving can only be performed on empty bags.
NovaSeptum® Sterile Sampling Units and Transfer System

SyringeBag Transfer SystemTube Transfer System
Sampling Unit Volume5 mL and 20 mL250 mLNot applicable
Maximum Pressure Conditions (at 25°C) 0.50 bar (7.25 psi)
Temperature RangeSingle 5ml : -80 to 134 °C (-112 to 273 °F)
Single 20ml : -80 to 121 °C (-112 to 250 °F)

Multi: -20 to 50°C (-4 to 122 °F)
-80° (When used with a NovaSeptum Case) to 50 °C (-112 to 122 °F)Polyethylene: -20 to 50 °C (-4 to 122 °F)
Silicone : -50 to 95 °C (-58 to 203 °F)
Materials of Construction
Trigger
Septum:Medical Grade platinum-cured silicone
Body:Polyester
Cannula:ASTM 316L Stainless steel
Sampling Bag:Not ApplicablePolyethylene film (PureFlex film)Not applicable
Fluid Contact Layer:Polycarbonate, medical grade platinum-cured silicone, medical silicone fluidPolyethylene film (PureFlex)Not applicable
Tubing:SiliconePolyethylenePolyethylene or Silicone
Drainage Bag (valid for Multi-Sampling Application):Polyethylene film (PureFlex film)Not ApplicableNot Applicable
Fittings
Inlet Tubing:Septum with a 1 mm or a 2 mm needle 
Outlet Tubing:Tubing: male and female luerSeptum Unit with a 1 mm or 2 mm needle3-piece Luer, containing a male, female and an injection site
Autoclaving*Single : autoclavable Multi : Not autoclavableNot autoclavableNot Autoclavable
USP <88> Class VIAll components in contact with sampling liquid conform with USP <88> Class VI.
Integrity TestingUnits are integrity tested at regular intervals during manufacturing.
AssemblyAssembled under ISO Clean Room Class 8 conditions in a facility certified to ISO 14644-1
SterilizationThe sampling units are sterilized by beta irradiation at ≥ 25 kGy  according to ISO 11137
Bacterial EndotoxinAqueous extraction contains
* Autoclaving can only be performed on empty bags.
NovaSeptum® Case
Compatible with Sampling Unit Volume50 mL, 100 mL and 250 mL
Maximum Pressure Conditions (at 25°C)0.50 bar (7.25 psi) up to 250 mL
Operating Temperature-80 to 37 °C (-112 to 98.6 °F)
Materials of ConstructionAPET (Amorphous Polyethylene Terephthalate) noble
SterilizationNon-sterile
NovaSeptum® Single-Use holder, TC 1.5 / 2“, 2 mm
Holder and MagazinePEI Black
Locking ringPEI Orange
ScrewA2-A4
Environmental
Design Temperature, TC holderSteam in place cycle at up to 134 °C (273 °F) or Autoclave during 1 hour at up to 134 °C (273 °F) (when applicable)
TraceabilityThe product and packaging label includes the catalogue number, lot number and expiration date.
SterilizationGamma irradiation between 25 kGy and 40 kGy
Component Material ToxicityAll wetted components comply with USP USP <88> Class VI.
Endotoxin level< 2.15 EU/device for all wetted components
AssemblyThis product is manufactured in a clean room environment complying with class 8 according to ISO® 14644-1.
PackagingEach Single use pre-loaded holder is double packed.