SPECIFICATIONS:

PINNACLE PCX SPECIFICATIONS

  • Dimensions: (h x w x d): 54 x 26.7 x 46.4 cm

  • (21.25 x 10.5 x 18.25 inches)

  • Weight: 54.4 lbs (24.7 Kg) – Duplex

  • Electrical: 100-120 V, 50/60 Hz, 1.7 A, 200 W or 200-240 V, 50/60 Hz, 0.8 A, 200 W

  • Mains voltage ± 10 % of nominal

  • Installation (over voltage) category II, pollution degree 2

  • Indoor use only

  • Environmental: Altitude up to 6500 ft

  • Ambient Temperature 15 °- 25 °C

  • Relative humidity up to 80 % at 31 °C

REAGENT PUMPS

  • True pulse-free syringe pump

  • Single piece ceramic barrel

  • Completely inert flow path

  • Maximum operating pressure 500 psi

  • Programmable flow rate

  • Flow range: 50 μL to 1500 μL/minute

  • Flow Accuracy 3 %

  • Flow Precision 0.5 % RSD

  • Refill cycle of 60 seconds

  • Automatic piston wash

  • Automatic reagent flush cycle

  • No check valves

CONTROL

  • Computer controlled

  • Windows PC with XP, Vista or Windows 7

  • USB connection or network

REACTOR

  • Heated reactor for temperature from 5 °C above ambient to 130 °C

  • Easy replacement reactor coil cartridges

  • Range of reactor dwell volumes; 0.1 mL to 3.5 mL

  • Reaction coil withstands up to 42 bar (600 psi) inlet pressure at 130 °C

  • Thermal safety switch limits temperature to 150 °C to prevent damage

  • Fast response

SAFEGUARDS

  • In-line check valve: Prevents reagent back flow into the column when HPLC pressure drops

  • Post-column system over pressure: A pre-calibrated relief valve opens at 35 bar (500 psi) to prevent rupture of the post-column reactor tubing in the event of down-stream blockage

  • Back-pressure regulator: Applies 7 bar (100 psi) to the detector flow cell outlet (waste) to prevent detector noise and precipitation due to out-gassing or boiling

COLUMN HEATER AND REACTOR CONTROLLER

  • Heater accepts up to 6 or 8 mm OD (0.25 or 0.31 inch) x 50-250 mm in length (column and guard)

  • Programmable temperature gradient

  • Power cooling capability

  • Temperature holds within ± 0.4 °C from the set point. Can be set with 1 °C resolution from 5 °C above ambient to 75 °C

  • Easy column access

INSTRUMENT PACKAGE AND FLOW PATH

  • Advanced fluidics valve management system

  • Easy access to internal components

  • Standard fittings

  • Post-column pressure relief valve

  • Side panels easily removed for service access

  • Integrated reagent reservoir tray

  • Corrosion proof pan and panels

DISPLAY

  • Backlit LCD

  • Real time temperature and pressure display

  • System status icons

  • Simple system control interface

GAS PRESSURE MANIFOLD AND REGULATOR

  • Panel mount manifold

  • Regulator maintains 0.3 bar (3-5 psi) on reagent reservoirs with 3-5 bar (45-75 psi) source pressure

  • Safety pressure relief valve opens at 1 bar (14 psi)

  • Manifold has two 1/4-28 tubing connections

  • Gas lines with anti-syphon valves

PRESSURIZED REAGENT RESERVOIR

  • One liter capacity (2 L reservoirs available)

  • Maintained under inert gas pressure to inhibit oxidation of oxygen-sensitive reagents

  • Valve built into reservoir cap permits sparging during reagent preparation

  • Reagent reservoirs fitted with 3.1 mm (1/8”) OD, oxygen-impermeable Air Barrier tubing for oxygen-sensitive reagents