Measurement type: Affinity (KD)
Measurement type: Enthalpy ∆H
Measurement type: Entropy ∆S
Measurement type: Stoichiometry (n)
Sample volume: 280 µL
Cell volume: 200 µL
Injection syringe volume: 40 µL
Injection volume precision: < 1% @ 2 µL
Sample throughput: 0-12 per 8 h day
Cell material: Hastelloy
Cell: Coin-shaped
Noise: 0.15 ncal/s
Temperature range: 2°C to 80°C
Temperature stability: ± 0.00012°C
Response time: 8 s*
Multiple feedback modes: Yes (passive, high gain, low gain)
Notes: *The MicroCal PEAQ-ITC Instrument Response Time is a true time constant. It is the time interval between the first deviation away from the baseline, and the point on the peak that is 63% of the maximum peak height.
Temperature (operating): 10°C to 28°C
Humidity: 0% to 70% RH, non condensing
Power: 100 - 200 V, 50/60 Hz, 130 W
Weight: 13.6 kg
Dimensions (W, D, H): 43 x 38 x 46 cm (calorimeter + wash station)