美国CST 氯分析仪
参数指标我要纠错



 

General


Range

Free or Combined Chlorine, HOCL, OCLˉ, Chloramines, and Chlorine Dioxide in Water or Chlorine Gas:  PPM to % Levels

Repeatability

± 1% of Full Scale

Measurement Principle

UV Absorbance

Light Source

Xenon Flash Lamp

Detector

Silicon Photodiode

Fiber Optic Cables

(2) 2 meter, 600 micron core

Sample Introduction

Titanium Cross Flow Cell or  PVC Long Pathlength Flow Cell

Process Pressure

2000psi max

Minimum Flow Rate

100 ml/min

Calibration

Analyzer is calibrated with customer sample; measurement normalized by zeroing every 1-2 months or as needed.

Path Length

Application dependent

Response Time

1 second

Power Requirement

24VDC nominal (12-48VDC), 8.5 watts max

Dimensions of Photometer

2.8” H x 4.5” W x 9.5” L

Weight of Photometer

~3.5 lbs. (1.6 kg)

Enclosure

NEMA4X anodized aluminum

PX2+ Operating Conditions


Process Temperature

204°C

Operating Temperature

5°C to 50°C

Storage Temparture

-20°C to 50°C

Communications


Outputs

4-20mA, RS-485 (MODBUS), or USB

Alarms

Contact closure (60VDC, 0.75 A max)

Display

3.2” capacitive touch LCD

 

 



 

 

行业:

?农业

?塑料

?纸浆和纸张

?纺织品

?废物处理

 

应用:

?漂白生产

?冷却水

?工业废水

?饮用水

?氯气