主要技术参数:
控温范围: 室温---450 ℃
控温精度: ±0.2 ℃
显示分辨率: 0.1 ℃
最大功耗: ≤500 W
温度恢复时间: 4 min
活塞位置检测:(MVR)
上下环距离: 30 mm
位移控制精度:±0.1 mm
时间精度:±0.15S
测试范围:0.03-450g/10min
标准件参数:
料筒参数:内孔 Φ9.55±0.025mm
活塞参数:活塞头Φ=9.475±0.015mm;
活塞头长度 H=6.35±0.1mm
口膜参数:挤出孔Φ=2.095±0.005mm
口模配置:Φ2.095±0.005/Φ1.18±0.01mm
砝码参数:
砝码精度:±0.5%
基本配置: A 0.325 KG(含压料杆)
B 0.875 KG
C 0.960 KG
D 1.200 KG
E 1.640 KG
注:可根据用户要求加配备砝码
五 主要配置及附件:
配 件 | 数 量 | 备 注 |
主 机 | 一台 | |
微型打印机 | 一台 | |
砝 码 | 一套 | 0.325kg、0.875kg、0.960kg、1.200kg、1.640kg(可加配) |
口 膜 | 一只 | Φ2.095±0.005mm |
压料杆 | 一只 | |
清料杆 | 一套 | |
加料漏斗 | 一只 | |
口膜通径杆 | 一只 | |
接料盘 | 一个 | |
连接电缆 | 一套 | |
主机地脚 | 四只 | |
附件箱 | 一个 | |
说明书 | 一本 |
Product Name: melt flow rate measurer
Two type: RTSL-400C
The three main purposes and functions:
Melt index apparatus complies with the requirements of the GB/T3682-2000 standard, the main parameters of melt index instrument but also meet the requirements of ISO1133-97 standard,ASTM1238 can be used to be used for melt flow rate of polyethylene, polypropylene,polystyrene, ABS, polyamide fiber, acrylic, resin, polyoxymethylene, fluorine plastic,polycarbonate and other plastic materials (MFR) or the melt volume flow rate (MVR) for the determination of;
The characteristics of melt index instrument:
1, the instrument adopts the micro processing control technology, digital control and integration.According to standard test process procedures setting.
2, fast heating speed, high precision of constant temperature; in the packing after, can quickly return to a state of constant temperature.
3, Chinese color crystal display, voice prompts, touch screen operation, easy to understand.
4, using imported double display temperature control, high control precision, temperature control precision and stability.
5, automatic cutting, cut hard plastics.
6, can be equipped with displacement measuring device (B, Guang Ou) is relatively large for measuring index volume method of material;
7, micro printer for printing data.
8, the heating element has eliminated the instrument common electric heating wire similar products used in the way, and the use of exotic alloys nichrome metal die casting, has thecharacteristics of long time use of state of high temperature burn not bad.
New features:
1, the instrument is upgraded to a touch screen control, style novel, convenient operation and convenient, more humane.
2, the replacement of the original monochrome liquid crystal display, using 65535 true colorscreen size, better visual effect, interface display more information.
3, improve the test results of function, to retain the decimal, the result four to five homes, which makes the calculation results more accurate.
4, setting the parameters can be set arbitrarily, easier to test personnel can meet the requirements of operating flexibility, more test.
5, the test of any step can be manually skip, convenient test.
6, the test results can be stored, and can be permanently stored data are poured out to U diskof computer, which is convenient for future check each batch of sample results, and compares the.
7, the instrument can be used for upgrading data by U disk or remote, file upgrade procedures,functions more powerful.
8, this instrument has the function of computer communication interface, can be tested bymeasurement with a program of software, more advanced technology.
(according to the special requirements of users with English operation method, also can be equipped with volume method).
Four main technical parameters:
Room temperature ---450 DEG C temperature range:
Temperature control precision: - 0.2 DEG C
Display resolution: 0.1 C
The maximum power consumption: less than 500 W
Temperature recovery time: 4 min
The piston position detecting: (MVR)
The upper and lower ring distance: 30 mm
The control precision of the displacement: + 0.1 mm
The time accuracy: + 0.15S
Test range: 0.03-450g/10min
Parameters of standard parts:
The barrel inner hole parameters: Phi 9.55 + 0.025mm
The parameters of piston: piston head phi = 9.475 + 0.015mm;
The piston head length H = 6.35 + 0.1mm
Export membrane parameters: the extrusion hole phi = 2.095 + 0.005mm
Die configuration: Phi 2.095 Phi 1.18 + 0.01mm + 0.005/
The weight parameter:
The weight accuracy: + 0.5%
The basic configuration: A 0.325 KG (including pressing rod)
B 0.875 KG
C 0.960 KG
D 1.200 KG
E 1.640 KG
Note: according to user requirements and equipped with weights
Five main configuration and accessories:
Fitting
The number of
Note
The main machine
A
Micro printer
A
A weight
A set
0.325kg, 0.875kg, 0.960kg, 1.200kg, 1.640kg (to be added)
Oral membrane
A
Phi 2.095 + 0.005mm
Pressing rod
A
Discharging rod
A set
Charging hopper
A
Oral membrane diameter rod
A
The material receiving disc
A
Connect cable
A set
Host anchor
Four
The accessory box
A
Manual
A Book