TiF-15SP-12 | TiF-15SP-15 | TiF-15SP-30 | TiF-15SP-50 | TiF-15SP-TU | |
Compressed pulse duration with external compressor1),5) | <12 fs | <15 fs | <30 fs | <50 fs | 15-50 fs (tunable) |
Spectrum width (FWHM)1) | >80 nm | >60 nm | >30 nm | >18 nm | 18-65 nm |
Central wavelength tuning | 800±15 nm (fixed) | 790-820 nm | 770-840 nm | 750-860 nm | 790-820 nm (FWHM >65 nm) 750-860 nm (FWHM >18 nm) |
Output power2) | >230 mW (3W pump) | >280 mW (3W pump) | >350 mW (3W pump) >470 mW (4W pump) | >400 mW (3W pump) >530 mW (4W pump) | >300 mW@65 nm (3W pump) >530 mW@18 nm (4W pump) |
General optical specifications | |
Pulse repetition rate (fixed) | 80±10 MHz |
Pump laser | integrated or stand-alone DPSS 532 nm low-noise laser (up to 4 W) |
Beam mode | TEMoo |
Beam diameter (at 1/e^2) | <2 mm |
Output polarization | linear, horizontal, PER >20 dB |
Beam divergence | <1 mrad |
Long-term stability2) | <0.1 % rms |
Noise4) | <0.2 % rms |
Physical dimensions (L × W × H) | |
Laser head dimensions | 513 × 268 × 120 mm (including integrated pump laser) |
Pump laser control unit dimensions | 353 × 360 × 119 mm |
Closed-loop chiller dimensions | 430 × 340 × 190 mm |
Environmental and utility specifications | |
Operating temperature | 15-30 °C |
Relative humidity | <60%, non-condensing |
Voltage | single-phase; 100-240 VAC; 50/60 Hz |
Power consumption | <2 kW |
Available configuration packages3) | |
Thermally stabilized monolithic body | included in any package |
Pulse train SMA output with mode-lock LED | included in any package |
“Auto” factory package | - built-in spectrometer - single-touch wavelength tuning w. presets - built-in power meter - active output power stability locking - Windows software with configurable widgets - automatic mode-lock start and monitoring PC requirements: USB 2.0 port, Windows 10 |
“Basic“ factory package (default) | - mode-lock LED and push-button starter - USB 2.0 wavelength tuning via step-motor slit (via step number information and calibration) PC requirements: USB 2.0 port, Windows 10 |
“Manual” factory package | - push-button starter - manual wavelength tuning slit |
1) - when tuned to 800 nm central wavelength; 2) - after 30 min warm-up with cold start, during 12-hour continuous operation under equal room temperature conditions using recommended stabilized closed-loop chiller with proper capacity and recommended low-noise on-board integrated highly stable pump laser with active power locking; 3) - please select one of the packages for your system; certain features may be tailored or combined differently according to specific customer requirements; 4) - measured from 10 Hz to 10 MHz; 5) - external compressor is not included. |