钛宝石飞秒激光器 TiF...
参数指标我要纠错


Standard models and available pump laser power matrix




TiF-100ST (standard wavelength range)TiF-100LW (extended IR range)

Pump laser powerWavelength tuningOutput powerWavelength tuningOutput power
3 W740-860 nm>350 mW аt 800 nmn/an/a
4 W735-870 nm>500 mW аt 800 nmn/an/a
5 W730-875 nm>650 mW аt 800 nm830-950 nm>200 mW at 950 nm
6 W725-880 nm>850 mW аt 800 nm830-960 nm>200 mW at 960 nm
8 W720-885 nm>1.25 W аt 800 nm830-970 nm>200 mW at 970 nm
10 W715-890 nm>1.55 W аt 800 nm830-980 nm>150 mW at 980 nm
It is possible to combine both ranges in a single laser system, please enquire for details.



General optical specifications
Output pulse duration1)<100 fs
Spectrum width1) (FWHM)>8 nm
Pulse repetition rate (fixed)80±10 MHz
Pump laserintegrated or stand-alone DPSS 532 nm low-noise laser (up to 10 W)
Beam modeTEMoo (M^2 <1.2)
Beam diameter (at 1/e^2)<2 mm
Output polarizationlinear, horizontal, PER >20 dB
Beam divergence<1 mrad
Long-term stability2)<0.1% rms
Noise3)<0.2% rms
Physical dimensions (L × W × H)
Laser head516 × 350 × 118 mm (including integrated pump laser)
Pump laser control unit353 × 360 × 119 mm
Closed-loop chiller430 × 340 × 190 mm
Environmental and utility specifications
Operating temperature15-30 °C
Relative humidity<60%, non-condensing
Voltagesingle-phase; 100-240 VAC; 50/60 Hz
Power consumption<2 kW
Available configuration packages4)
Thermally stabilized monolithic bodyincluded in any package
Pulse train SMA output with mode-lock LEDincluded in any package
Auto” factory package- built-in spectrometer
- single-touch wavelength tuning w. presets
- built-in power meter
- active output power stability locking
- Windows software with configurable widgets
- automatic mode-lock start and monitoring
PC requirements: USB 2.0 port, Windows 10
“Basic“ factory package (default)- push-button starter
- USB 2.0 wavelength tuning via step-motor slit (via step number information and calibration)
PC requirements: USB 2.0 port, Windows 10
Manual” factory package- push-button starter
- manual wavelength tuning slit
1) - when tuned to 800 nm central wavelength;
2) - after 30 min warm-up with cold start, during 12-hour continuous operation under equal room temperature conditions using recommended stabilized closed-loop chiller with proper capacity and recommended low-noise on-board integrated highly stable pump laser with active power locking;
3) - measured from 10 Hz to 10 MHz;
4) - please select one of the packages for your system; certain features may be tailored or combined differently according to specific customer requirements.