TETA-6 | TETA-10 | TETA-20 | |
Maximum average output power | 6 W | 10 W | 20 W |
Maximum pulse energy | >400 uJ | >1 mJ | |
Minimum pulse duration1) | <270 fs | <350 fs | |
Pulse duration tuning range1) | <270 fs - 10 ps | <350 fs - 10 ps | |
Central wavelength2) (fixed) | 1033±3 nm | 1030±3 nm | |
M^2 | <1.15 | <1.25 | |
Beam diameter (at 1/e^2) | 3±0.3 mm | 5±0.5 mm | |
Pulse repetition rate (user-adjustable) | single-shot…200 kHz (up to 1 MHz upon request) | ||
Pre- and post-pulse contrast | >5000:1 | ||
Long-term output stability3) | <0.5% rms over 48 h | ||
Output polarization | linear, vertical | ||
Spatial mode | TEM00 | ||
Beam ellipticity4) | <10% | ||
Beam astigmatism4) | <10% | ||
Beam divergence (full angle) | <0.6 mrad | ||
Beam pointing stability | <25 urad/°C | ||
Cold start warm-up time (beam position, output power) | <40 min | ||
Environmental and utility specifications | |||
Operating temperature | 15-30 °C | ||
Relative humidity | <60%, non-condensing | ||
Voltage | single-phase; 100-240 VAC; 50/60 Hz | ||
Power consumption | <1.5 kW | <2 kW | |
Physical dimensions | |||
Laser head (LxWxH) | 460x250x147 mm | 500x330x147 mm | |
Control and power supply 19“ rack (WxLxH) | 553x600x663 mm | ||
Umbilical length | 3 m | ||
1) - measured with Avesta's AA-10DD-12PS interferometric autocorrelator using Gaussian fitting; motorized tunable pulse duration with PC control up to 10 ps is also installed by default; the Compulse-1030 external hollow-fiber pulse compressor with output pulse duration down to 30 fs is also available as an option; 2) - external second, third and fourth harmonic generators are available upon request; certain built-in combinations are also available; Raman shifters to 1530 nm and 1890 nm are also available; 3) - measured under stable environmental conditions; 4) - measured at maximum average power and 100 kHz repetition rate. |