Biotage TurboVAP LV...

Biotage TurboVAP LV新款全自动氮吹浓缩仪LV型参数指标

参数指标我要纠错

供气压力:6 bar (0.6 MPa, 87 PSI)

供气流速:48/24位 200/160 升/min

水浴温度:室温-90℃,准确度±2℃

升温速度:20-55℃,40分钟

气流范围:48/24位  3.5/5.5升/分钟




Technical Specifications

Evaporation

Gas Supply and Connection

The system can operate using inert gas. Compressed air can be used but is not recommended with some solvents due to the possible risk of explosion. The gas tube supplied with the system has a 6 mm outer diameter. An adapter for connecting a gas tube with an outer diameter of 1/4“ can be purchased separately.

Note: There is no regulator for the gas inlet supplied with the system. This can be purchased separately as an optional accessory (P/N 352281SP Air Control Unit and P/N 353480SP Supply Connection Set).

  • Required supply pressure for full flow capability全流速供气压力: 6 bar (0.6 MPa, 87 PSI).

  • Maximum pressurezei大压力: 9 bar (0.9 MPa, 130 PSI).

  • Required supply flow for full flow capability for each manifold: 供气流速:48/24位 200/160 升/min

    • 48 positions: 200 litres/min

    • 24 positions: 160 litres/min

Water Bath Temperature Range 水浴温度范围

Ambient to 90°C, ± 2°C accuracy  室温-90℃,准确度±2℃

Temperature Increase 升温速度

Heating time from 20°C to 55°C is approx. 40 mins.  20-55℃,40分钟

Gas Flow Range 气流范围

The maximum gas flow depends on the gas supply pressure and flow, the number of available nozzles and how many that are open. The maximum gas flow for each nozzle:

  • 48 positions: 3.5 litres/min

  • 24 positions: 5.5 litres/min                 

Chemical Resistance 抗化学试剂能力

  • Trifluoroacetic acid 5 %  三氟乙酸5%

  • Trichloroacetic acid 5 % 三氯乙酸5%

  • Formic acid 5 %  甲酸5%

  • Acetic acid 10 % 醋酸10%

  • Phosphoric acid 5 % 磷酸5%

  • Propionic acid 5 % 丙酸5%

  • Ammonium hydroxide 10 % NH4OH10%

  • Triethylamine 10 %   三乙胺10%


Racks and Tube Sizes 管架

TurboVap LV multi racks

  • 48 positions for 10-16 mm (OD) x<45 mm (length) mini vials

  • 48 positions for 10-20 mm (OD) x 75-165 mm (length) tubes

  • 24 positions for 12-30 mm (OD) x 75-165 mm (length) tubes


Interfaces界面

Touch Screen  触摸屏  

7 Inch Touch Screen       7英寸触摸屏

Auxiliary

Connection for end-point detection      

System

Fume Hood/Ventilation

The system must be either connected to ventilation system using the exhaust outlet at the rear of the system or placed in a well-ventilated fume hood or an equivalent enclosure to reduce the risk of exposure to harmful solvent vapours. The ventilation system or fume hood must be capable of exhausting 1 m3/min (35.3 ft3/min).

Connection: The exhaust hose supplied with the system is 3.8 meters (12.5 feet) and has a 50.8 mm (2“) inner diameter.

Operating Temperature

18°C to 32°C

Storage Temperature

-25°C to 60°C

Humidity

20% to 90% RH (non-condensing)

Electrical Supply

100-127 V and 220-240 V at 50/60 Hz

Fuses

F10A at the power inlet (1 required)

Max Power Consumption

1000 VA

Weight

17 kg (37.5 lbs) when the water bath is empty and maximum 29 kg (63.9 lbs) with deionized water, rack, and tubes.

Dimensions (W x D X H)

400 x 400 x 390 mm (15.7“ x 15.7“ x 15.4“) including the hinges at the side and exhaust at the rear. The height with the lid open is 590 mm (23.2“).

Max Sound Level

75 dB(A)

Certifications

CE marked and CB and NRTL certified.

  气源要求 zei小入口压力:60psi;

  zei大入口压力:100psi;

  气体排空埠和管道 随机提供5.1cm 内径;3.8m 长排空管;

  电源要求: 230V,50Hz;

  尺寸: 54cm 宽×30cm 高×31cm 深。