Baseline 9000 MNME甲...
参数指标我要纠错


● FID检测器,自动点火,自动切断样品气,燃气和助燃空气
● FlowGuard电子控制燃气,空气和样品气
● 根据用户定义浓度自动校准
● 采用样品旁路系统电子反压调节,确保快速响应
● 浓度&错误报警功能
● 用于浓度报警,事件和诊断的可编译继电器
● 通过RS232和以太网远程控制
● 图形化显示,用户友好简单操作系统



Analytical Specifications

Detector

Flame ionization detector (FID)

Oxidizer

Oxidizes NonMethane gasses

Range (as Methane)

MDQ 0.03 ppm, Full-scale 1 - 200 ppm (O2 Balance)MDQ 0.06 ppm, Full-scale 1 - 200 ppm (Air Balance)MDQ 0.1 ppm, Full-scale 1 - 1,000 ppm (Air Balance)

Repeatability

±1% Full-scale response

Drift

±1% of Full-Scale over 24 hours

Response Time

T90 < 30 sec, Switching < 3 minutes (The Sample Mode switching time is programmable)

Alarms

Multilevel concentration and fault alarms that result in an audible and visually displayed alarm. 
Alarms may also be mapped to relays to control external equipment.

Sampling

Internal single or multipoint modules for pre-filtered (1 micron), non-condensing samples, with or without sample pump.

Calibration

Programmable automatic or manual calibration

Support Gas

Hydrogen 35 cc/min (H2He Blend 100 cc/ min), Air 175 cc/min (typical). Hydrocarbon content must be less than 1 ppm. Fuel blend options available, consult Baseline®.


Physical Specifications

Dimensions

19.00“ (48.3 cm) W x 14.25“ (36.2 cm) D x 5.25“ (13.3 cm) H

Weight

< 20 lb (9.07 kg)

Configuration

Bench-top or 19” (48.3 cm) rack-mount, 3U

Connections

1/4“ (6.35 mm) tube fitting connectors

Operating Temperature

32 to 104°F (0 to 40°C)

Operating Humidity

0 - 95% (non-condensing)


Electrical Specifications

Power

100–230 V AC, 50/60 Hz, 2 A

Display

Graphical LCD display, 3.4” x 4.5” (8.64 x 11.43 cm)

Relay Outputs

5 programmable form A relays rated to 3A @ 230V AC (optional additional 9 relay output board)

Analog Outputs

1 programmable 0-20mA or 4-20mA isolated output (optional additional 3 analog output board)

Digital Outputs

RS-232, Ethernet


相关VOC检测仪/TVOC监测仪