zei宽的分子量检测范围:10-240kD
12次CE-SDS检测仅需5-7小时
FEATURE | CE-SDS SPECIFICATION |
---|---|
Minimum Sample Volumezei小样品体积 | 50 µL |
Sample Delivery样品传输 | Electrokinetic电动力 |
Typical Separation Time典型分离时间 | Reduced 还原IgG: 25 min, Non-reduced 未还原IgG: 35 min |
Detection Capability检测能力 | UV Absorbance at 220 nm 紫外吸收220nm |
Typical Voltage分离电压 | Separation: 5750 V |
Sample Injections per Cartridge每卡盒进样数 | 100 guaranteed, 200 maximum 确保100次,zei多200次 |
Maximum Sample Injections per Batch每批zei大进样 | 48 |
Size Range 分子量检测范围 | 10–240 kDa |
Sizing CV 检测偏差 | ≤2% |
Relative Migration Time CV 相对位移时间偏差 | <1% for reduced IgG |
Sizing Resolution 测量分辨率 | ≥1.5 for NGHC/HC IgG Standard |
Dynamic Range 动态范围 | 2 logs |
Linearity 线性 | >0.995 |
Sensitivity (LOD) 灵敏度(LOD) | 0.3 µg/mL (Value based on Internal Standard) |
Sample Tray Options 样品盘选项 | 96-well plates or 48 vials |