分析测试百科网

搜索

喜欢作者

微信支付微信支付
×

How to interrupt scintillation?

2019.5.20
头像

zhaochenxu

致力于为分析测试行业奉献终身

Peter Novick Lab, Department of Cell Biology Yale University School of Medicine HOW TO PERFORM AN INTERRUPT OF AN
ONGOING AUTOCOUNT
1). This program interrupts the ongoing autocount to allow the counting of a
small number of vials- only one rack of vials. If you have more than one
rack to count, think creatively and I'm sure you'll figure out how to count
them all with the help of the interrupt function.
2.) HOW TO USE THE INTERRUPT FUNCTION:
a). Load your samples onto a rack bearing the INTERRUPT card.
Put the interrupt card in the frontmost card holder (left side).
b.) Put your rack in with the other racks from the autocount. Place your
rack near the front (the scintillation counter will find the interrupt rack
no matter where you place it).
c.) Press the yellow INTR button. You are presented with the following
display printed on the paper where the counts are printed, and the
editing light will go on:
d). Most people will press -2- (INTERRUPT AUTO) and enter this by
pressing the RETURN button, since most people will have a rack of
samples to count rather than a single tube (-1-).
e). Press blue USER NUMBER button (don't press the return button). The
scintillation counter then displays the various numbers of the counting
programs.
f). Press the number of the desired program for counting the interrupt
samples and press the STORE PROG button.
g). Press MAN COUNT button.
h). The machine will automatically count your tubes and will resume the
previous Autocount after processing the last sample in the interrupt
rack. Your counts will be printed out on a peice of paper separate from
the previous autocount.
i). Be absolutely certain when finished with the interrupt function that
the previous autocount has been continued!!!
j). More information about this function is found in the manual for the
scintillation counter on page 4-2.
 