IEC 60904-7-2008
光电器件.第7部分:光电器件的测量用光谱错配修正的计算

Photovoltaic devices - Part 7: Computation of the spectral mismatch correction for measurements of photovoltaic devices


哪些标准引用了IEC 60904-7-2008

 

BS EN 60904-5-2011光伏器件.用开路电压法测定光伏(PV)器件的等效电池温度(ECT)DIN EN 60904-2-2008光伏器件.第2部分:基准太阳能装置的要求(IEC 60904-2-2007)BS EN 60904-3-2008光电器件.用参考光谱照射数据对陆地光电(PV)太阳能装置的测量原理DIN EN 61646-2009薄膜地面光电(PV)模块.设计鉴定和定型BS EN 61646-2008薄膜地面光电(PV)模数.设计合格鉴定和型号批准DIN EN 60904-3-2009光电器件.第3部分:带基准光谱辐照数据的地球光电(PV)太阳能器件的测量原理BS EN 60904-1-2006光电器件.光电电流-电压特性的测量IEC 61646-2008薄膜地面光电(PV)模块.设计鉴定和定型DIN EN 60904-1-2007光电器件.第1部分:光电器件电流-电压特征测量(IEC 60904-1:2006)BS EN 60904-2-2007光电器件.基准太阳能装置的要求BS EN 60904-9-2007光伏器件.第9部分:太阳模拟器性能要求IEC 60904-3-2008光伏器件.第3部分:具有标准光谱辐照度数据的地面用太阳光伏(PV)器件的测量原理IEC/TS 62941-2016IEC 61829-2015光伏 (PV) 阵列. 电流电压特性的现场测量IEC 60904-1-1-2017BS EN 60904-2-2015光电器件.对光伏基准装置的要求IEC 60904-8-2014光伏器件 第8部分:光伏器件光谱响应度的测量BS EN 60904-8-2014光电器件.光电器件波谱反应测量BS EN 60904-3-2016光电器件.用参考光谱照射数据对陆地光电(PV)太阳能装置的测量原理IEC/TS 62941-2016地面光伏(PV)模组. 光伏模块设计评定和型式批准信心增加指南DIN EN 61829-2016光伏(PV)阵列.电流电压特性现场测量(IEC 61829-2015).德文版本EN 61829-2016DIN EN 60904-1-1-2018光伏设备.第1-1部分:多接点光伏(PV)设备电流电压特性的测量(IEC 60904-1-1-2017); 德文版本EN 60904-1-1-2017DIN EN 60904-8-2015光伏器件.第8部分:光伏器件光谱响应的测量(IEC 60904-8-2014);德文版本EN 60904-8-2014IEC 60904-1-1-2017光伏设备. 第1-1部分: 多接点光伏(PV)设备电流电压特性的测量BS EN 61215-2-2016BS PD IEC/TS 61836-2016太阳光伏能源系统.术语,定义和符号BS PD IEC/TS 62941-2016地面光伏 (PV) 模块. 光伏模块设计资质和型式批准的置信度增加指南BS EN 60904-2-2015光电器件.对光伏基准装置的要求BS EN 61829-2016光伏 (PV) 阵列. 电流电压特性现场测量

IEC 60904-7-2008



标准号
IEC 60904-7-2008
发布日期
2008年11月
实施日期
废止日期
中国标准分类号
L50
国际标准分类号
27.160
发布单位
IX-IEC
引用标准
IEC 60891 IEC 60904-1 IEC 60904-2 IEC 60904-3 IEC 60904-8 IEC 60904-9 IEC 60904-10 IEC 61215 IEC 61646
被代替标准
IEC 82/540/FDIS-2008 IEC 60904-7-1998
适用范围
This part of IEC 60904 describes the procedure for correcting the bias error introduced in the testing of a photovoltaic device, caused by the mismatch between the test spectrum and the reference spectrum and by the mismatch between the spectral responses (SR) of the reference cell and of the test specimen. The procedure applies only to photovoltaic devices linear in SR as defined in IEC 60904-10. This procedure is valid for single junction devices but the principle may be extended to cover multijunction devices. The purpose of this standard is to give guidelines for the correction of measurement bias, should there be a mismatch between both the test spectrum and the reference spectrum and between the reference device SR and the test specimen SR. Since a PV device has a wavelength-dependent response, its performance is significantly affected by the spectral distribution of the incident radiation, which in natural sunlight varies with several factors such as location, weather, time of year, time of day, orientation of the receiving surface, etc., and with a simulator varies with its type and conditions. If the irradiance is measured with a thermopile-type radiometer (that is not spectrally selective) or with a reference solar cell, the spectral irradiance distribution of the incoming light must be known to make the necessary corrections to obtain the performance of the PV device under the reference solar spectral distribution defined in IEC 60904-3. If a reference PV device or a thermopile type detector is used to measure the irradiance then, following the procedure given in this standard, it is possible to calculate the spectral mismatch correction necessary to obtain the short-circuit current of the test PV device under the reference solar spectral irradiance distribution included in Table 1 of IEC 60904-3 or any other reference spectrum. If the reference PV device has the same relative spectral response as the test PV device then the reference device automatically takes into account deviations of the real light spectral distribution from the standard spectral distribution, and no further correction of spectral bias errors is necessary. In this case, location and weather conditions are not critical when the reference device method is used for outdoor performance measurements provided both reference cell and test PV device have the same relative spectral response. Also, for identical relative SR's, the spectral classification of the simulator is not critical for indoor measurements. If the performance of a PV device is measured using a known spectral irradiance distribution, its short-circuit current at any other spectral irradiance distribution can be computed using the spectral response of the PV test device.

IEC 60904-7-2008系列标准


谁引用了IEC 60904-7-2008 更多引用





Copyright ©2007-2022 ANTPEDIA, All Rights Reserved
京ICP备07018254号 京公网安备1101085018 电信与信息服务业务经营许可证:京ICP证110310号