The Carbohydrate Removal Cartridge (CRC) is an in-line sample pretreatment cartridge that removes carbohydrates from your amino acid sample.
In-line sample pretreatment cartridge eliminates laborious sample cleanup.
Removes carbohydrates from amino acid samples, ensuring reliable amino acid quantification.
AAA-Direct™is a simple new approach for separating and detecting amino acids without derivatization.With this technique, it is also possible to separate anddetect select carbohydrates simultaneously with amino acids in a single run. For applications where carbohydrates are present, but only the amino acids are of interest, the CRC can be installed to remove the carbohydrates on-line.
The CRC is an in-line pretreatment cartridge packed with cation-exchange resin.The amino acids are retained and the carbohydrates are washed to waste.After the carbohydrates have been removed from the cartridge, the CRC is switched in-line with theAAA-Directsystem and the AminoPac®PA10 column separates the amino acids.
The CRC is used in conjunction with the carbohydrate removal accessory (CRA).In addition to the CRC, the CRA consists of a syringe pump and six-port valve. To install the CRA, theAAA-Directsystem must be modified.The syringe pump is connected to the autosampler, whereas the GS50 pump and AminoPac PA 10 are redirected to the six-port switching valve.
Comparison of Chromatograms with and without the SugarTrap Installed.
Ordering Information | |
---|---|
Carbohydrate Removal Accessory Kit for ICS-3/5000 DP | 064418 |
Carbohydrate Removal Accessory Kit for ICS-3/5000 SP | 070510 |
Carbohydrate Removal Cartridge (CRC) for AAA-Direct | 068598 |
2nd Channel Kit, LC20/30, Right Slide, Rear Loading | 052322 |
Acclaim SST Guard Cartridge Holder V-1 | 059456 |